INSTALLATION • OPERATION • MAINTENANCE
Resistivity Model: ASIIR
Conductivity/TDS Model: ASIIC
Water Quality Instrumentation
Accuracy Reliability Simplicity
ACCURACY • RELIABILITY• SIMPLICITY
AQUASWITCH I
&
AQUASWITCH II
Operation
Manual
24 August 05
Vedere la pagina 0
1 2 3 4 5 6 ... 50 51

Sommario

Pagina 1 - AQUASWITCH II

INSTALLATION • OPERATION • MAINTENANCEResistivity Model: ASIIRConductivity/TDS Model: ASIICWater Quality InstrumentationAccuracy • Reliability • Sim

Pagina 2 - QUICK REFERENCE GUIDE!

The above is a Special Order High Temperature sensor with a cellconstant of 0.1, made of Titanium, and a Wet-Tap Valve.Additional Ordering Information

Pagina 3 - 21 may 03

RANGEResistivity0-20 MΩ0-10 MΩ0-5 MΩ0-2 MΩ0-1 MΩ0-500 KΩ0-200 KΩConductivity0-1 µS*0-1 ppm*0-2 µS*0-2 ppm*0-5 µS*0-5 ppm*0-10 µS0-10 ppm0-20 µS0-20 pp

Pagina 4 - AQUASWITCH I™

II. INSTALLATIONA. GENERALThis section provides the recommended procedures for properlyinstalling the AQUASWITCH I Controller, or AQUASWITCHII Monitor

Pagina 5 - TABLE OF CONTENTS

1. Select mounting site location. Mark and drill the requiredmounting holes. For hole drilling locations, see figureII.B.1.2. Insert the 1/4" X 2

Pagina 6 - TABLE OF CONTENTS Continued

2. Carefully unfasten and separate the AQUASWITCH'sfront panel from its enclosure.3. Disconnect all panel cable(s)/wires from theAQUASWITCH'

Pagina 7 - I. INTRODUCTION

A device to disconnect the AQUASWITCH from thepower supply is required. It is recommended thatthis switch or circuit breaker be labeled as thedisconne

Pagina 8 - DOUBLE INSULATED

The following procedures are to be used to install theAQUASWITCH II to 115 VAC main power source. For theprocedures to install the optional 230 VAC ma

Pagina 9

a. USP 25 MODIFICATION (No Temperature Compensation)This simple modification will allow your AQUASWITCH IIMonitor/controller to meet the USP 25 requi

Pagina 10

5. SOLID STATE OUTPUT 24 VDC Unregulated 30mA max. The following instructions areassuming the Monitor/controller enclosure is already open. a. Piezo

Pagina 11

6. SOLENOID VALVE CONNECTIONSThe solenoid valves require electrical power to operate. Thispower must be supplied by the user. See figure II.E.10 for t

Pagina 12 - II. INSTALLATION

AQUASWITCH I & II™QUICK REFERENCE GUIDE!If you read nothing else in thismanual please read this Quick Reference Guide. PLEASE READ and COMPREHEND

Pagina 13

7. ALARM/CONTROL RELAY CONNECTIONSThe Myron L Company AQUASWITCH is equipped with a “DryContact” relay which is designed to energize/de-energize whent

Pagina 14 - Figure II.D.2

F. 0-10VDC RECORDER OUTPUTThe 0-10VDC output is designed to give the user the capability ofsending a signal to a remote meter, recorder, PLC or SCADAs

Pagina 15 - Figure II.E.1

G. RE-RANGE YOUR AQUASWITCH II(Range Module Installation)1. DESCRIPTIONThe AQUASWITCH II Monitor/controllers have been designedfor easy field re-range

Pagina 16 - Figure II.E.3

III. OPTIONS & ACCESSORIESA. 4A/4AO MODULE (4-20mA OPTION)-4A 4-20mA Self/Remote-powered Isolated output module ordered with Monitor/controller.4A

Pagina 17 - BK WT RD GN NU R- R+

PhysicalIf the front panel has all ready been removed from the enclosureskip to #3. 1. Using a standard slot screwdriver remove the two (2) screws on

Pagina 18

4. If not, adjust the CAL control marked “4mA” until theDVM indicates 4mA, see figure III.A.5.5. Press the Full Scale Test Switch, the DVM shouldindic

Pagina 19 - TRANSFORMER

4. CONVERTING A CURRENT TO A VOLTAGECurrent measuring devices actually measure voltage, but have aninternal resistor as shown in figures III.A.1 &

Pagina 20 - Figure II.E.12

B. PIEZO ELECTRIC ALARM -PA Piezo Electric Alarm ordered with AQUASWITCH II.PAO Piezo Electric Alarm ordered separately forAQUASWITCH II.1. DESCRIPT

Pagina 21

Test1. Turn power ON. Depending on configuration, the Piezowill sound off when —. NOTE: If the sensor is connected, the solution value, set pointvalue

Pagina 22 - MYRON L

C. TEMPERATURE MODULE-TP Temperature Module ordered with Monitor/controller.TPO Temperature Module Kit ordered separately. Requires; CSXX-TP sensor,

Pagina 23 - ACCESSORIES

QUICK SET POINT ADJUSTMENT - See Section IV.B.2.The set point setting is based upon the user's particular water purity specifications or requirem

Pagina 24 - (+) SIGNAL OUT

9. While holding the front panel, align the display to theopening and at the same time, align the Temperature Module mounting holes to the front panel

Pagina 25

3. RECALIBRATIONElectronic Calibration Only. For System Calibration, see below.The Temperature Module was calibrated at the factory, however,if you wi

Pagina 26 - SO PO PISO PO PI

7. If not, adjust the calibration control marked “ZERO” (seefigure III.C.1 for location).8. Install SPAN resistor (1758.56Ω) across sensorterminals BL

Pagina 27 - WATER QUALITY

D. REMOTE ALARM - RA™ RA Ordered as an accessory, includes 8” harness withconnector, and two (2) wire nuts. 1. DESCRIPTIONThis remotely mounted AUDIBL

Pagina 28

alarm for approximately three minutes or until youimprove the water quality (readjust controller set point).If three minutes muting is fine for your a

Pagina 29 - BR BL OR - +

IV. OPERATING PROCEDURESThe front panel illustrations, switch and indicator operationaldescriptions have been provided to assist the user in identifyi

Pagina 30

!CAUTION - READ FOLLOWING CAREFULLYB. SETUP PROCEDURES These Setup procedures cover (1) converting the alarm circuit totrigger on a decreasing (cond

Pagina 31 - Figure III.C.5

1. Using the “SET POINT CHECK” switch, verify that thedisplay indicates the desired Set Point.NOTE: If the display does not indicate the correct Set P

Pagina 32

V. COMPONENT IDENTIFICATION, CALIBRATION AND PREVENTIVE CAREThis section provides detailed illustrations to assist the user inidentifying the AQUASWIT

Pagina 33 - + - 10

B. “PURGE CYCLE” CALIBRATION PROCEDUREThe Purge Cycle control adjustment (see figure V.A.1) determinesthe time span that the Bleed and Process solenoi

Pagina 34

1AQUASWITCH I™MYRON LCOMPANYBANK ABANK BWATER QUALITYBLEEDIN USE GOOD POORBANK SWITCH ALARM RESETAQUASWITCH II™Model ASIIR-11(A Digital Resis

Pagina 35 - IV. OPERATING PROCEDURES

!CAUTION - READ FOLLOWING CAREFULLYD. CALIBRATION PROCEDURESAQUASWITCH I does not require calibration. All Myron L AQUASWITCH II Monitor/controllers

Pagina 36 - RESISTIVITY

3. SENSOR SUBSTITUTE CALIBRATIONNIST traceable Sensor Substitutes are commonly use to verifyand calibrate Resistivity Monitor/controllers. Normally th

Pagina 37

the area of the AQUASWITCH II Monitor/controllers operation,i.e. for a conductivity AQUASWITCH IIMonitor/controllers that is operating between 15 and2

Pagina 38 - V. COMPONENT IDENTIFICATION

VI. OPTIONS & ACCESSORIESA. OPTIONS ORDERED WITH MONITOR/CONTROLLERPART # DESCRIPTION-4A 4-20mA Self-Powered Isolated Output Module (ASII only)-P

Pagina 39 - Figure V.A.2

C. STANDARD SOLUTIONS & BUFFERS1. CONDUCTIVITY/TDS STANDARD SOLUTIONSYour Conductivity/TDS Monitor/controller has been factorycertified to meet ce

Pagina 40

VII. REPLACEMENT PARTSPRICES AS OF July 1, 2001PRICES SUBJECT TO CHANGE — contact the Myron L Company for the latest prices. Prices are in US dollars.

Pagina 41 - OEM INSTALLATIONS ONLY

VIII. LIMITED WARRANTY All Myron L Company AQUASWITCH I Controllers & AQUASWITCH II Monitor/controllers and sensors have a Two YearLimited Warrant

Pagina 42

IX. GLOSSARY442™ An Internationally recognized “natural water”standard developed by the Myron L Company in1964. Algorithm A procedure for solving a m

Pagina 44

XI. ADDENDUMA. CONDUCTIVITY, TDS, RESISTIVITY and TEMPERATURE RELATIONSHIPS1. TEMPERATURE COMPENSATION (Tempco) of Aqueous SolutionsElectrical conduct

Pagina 45 - VII. REPLACEMENT PARTS

TABLE OF CONTENTSSECTION PAGEAQUASWITCH™ ILLUSTRATIONS (ASI & ASIIR-11) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 46 - VIII. LIMITED WARRANTY

XI. ADDENDUM Continued4. RESISTIVITYResistivity is the inverse of conductivity (the reverse is alsotrue), i.e. 18.2 MΩ = 0.0549 µM or µS, and 0.2 µM/

Pagina 47 - IX. GLOSSARY

Water Quality InstrumentationAccuracy • Reliability • SimplicityACCURACY • RELIABILITY • SIMPLICITYMYRON L COMPANY2450 Impala DriveCarlsbad, CA 92010-

Pagina 48 - X. NOTES

TABLE OF CONTENTS ContinuedSECTION PAGE3. Recalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 49 - XI. ADDENDUM

I. INTRODUCTIONThank you for selecting one of the Myron L Company’s newestAQUASWITCH Monitor/controllers. The new AQUASWITCH™series is based on input

Pagina 50 - XI. ADDENDUM Continued

AMBIENT TEMPERATURE RANGE:32°F (0°C) to 140°F (60°C)HOUSING CONSTRUCTIONFully gasketed heavy-duty ABS for splashproof and corrosion resistance. Rated

Pagina 51 - Water Quality Instrumentation

E. SENSORSAll sensors feature integral thermistor style temperature sensorsto ensure accurate, rapid and reliable automatic temperaturecompensation.